Dell 3100 Especificaciones Pagina 287

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 376
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 286
Magnetic Force Microscopy
Advanced MFM Operation
Rev. D Dimension 3100 Manual 267
Setpoint
For the most reproducible results, it is best to use a consistent setpoint. In LiftMode, the total tip-
sample distance h
tot
is the sum of the average tip-sample distance in TappingMode h
T
, and the lift
scan height h
lift
(see Figure 15.3b). In TappingMode, the average tip-sample distance h
T
is equal to
the oscillation amplitude, which is determined by the setpoint and the amplitude sensitivity of the
tip. MESPs typically have an amplitude sensitivity of approximately 25nm/V.
Large variations in setpoint can change the total tip-sample distance in LiftMode, sometimes with
visible results in the magnetic image. For this reason, reproducible results are most easily obtained
by using consistent setpoints. Note that a lift scan height of 0nm still gives a mean tip-sample
distance of h
T
in LiftMode.
Figure 15.3b Tip Height and Oscillation Amplitudes (TappingMode & LiftMode)
The relationship between setpoint voltage and oscillation amplitude is known as the sensitivity. Its
value can be determined with Force Calibration (see Chapter 13). For 225µm MFM cantilevers,
the sensitivity is typically in the range 20-25nm/V. The exact value depends on the position of the
laser spot on the cantilever, the cantilever geometry, and the particular instrument. A 1V Setpoint
typically corresponds to h
T
~14-18nm.
h
tot
=
h
T
+ h
lift
h
lift
Lift Pass
TappingMode Pass
h
T
Vista de pagina 286
1 2 ... 282 283 284 285 286 287 288 289 290 291 292 ... 375 376

Comentarios a estos manuales

Sin comentarios