Electric Techniques
Electric Force Microscopy
Rev. D Dimension 3100 Manual 285
Jumper Configurations With Extender Electronics
As shipped from the factory, systems with the Extender Electronics option should have an original
backplane jumper configuration as shown in Figure 16.2j. This configuration connects both the tip
and sample to the ground.
Figure 16.2j Normal Jumper Configuration with Extender Electronics Module
CAUTION: Turn off the NanoScope controller and unplug the power cable
from the microscope electronic box before attempting to adjust
jumper configurations.
ChuckGround/Bias
Analog 2 or
Unused
Auxiliary 2 Output
Ground
STM Signal
Indicates Jumpers
Gnd/OSC + DC signal
Unused
Ground
Ground
AFM Tip
Chuck
Sample
Tip