Dell 3100 Especificaciones Pagina 354

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 376
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 353
Calibration
Calibrating Z
334 Dimension 3100 Manual Rev. D
4. If the two peaks do not appear in the display, increase the Histogram filter cut off in the
Configure dialog box to 2-10nm.
5. Open the Configure dialog box by clicking the Configure button in the Depth dialog box
(see Figure 17.8e).
Figure 17.8e Z Calibration Configure Dialog Box
6. In the Data result dialog box located under the image on the display screen, check the Peak
to Peak value and record it for future reference.
7. Click Quit to exit the Depth dialog box.
Figure 17.8f Z Calibration Depth Dialog Box
Vista de pagina 353
1 2 ... 349 350 351 352 353 354 355 356 357 358 359 ... 375 376

Comentarios a estos manuales

Sin comentarios